background image
36
>
0
20
40
60
80
19859
1
2
3
4
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
min
Conditions for both columns:
Column: Kinetex 1.7 µm XB-C18
Dimensions: 50 x 2.1 mm
Guard Cartridge: SecurityGuard ULTRA C18 (ODS) 2.1 mm ID
Part No.: AJ0-8768
Mobile Phase: Acetonitrile / Water (50:50)
Flow Rate: 0.5 mL/min
Detection: UV @ 254 nm
Sample: 1. Uracil
2. Acetophenone
3. Toluene
4. Naphthalene
Cartridge with Holder
Cartridge
Holder
· Increases column lifetime of virtually all manufacturers' UHPLC columns
· Offers more reproducible chromatography
· For pressures up to 20,000 psi
Parameters
With SecurityGuard uLTRA
Without SecurityGuard uLTRA
Difference
Selectivity
1.36
1.35
-0.99%
Efficiency (Plates/Meter)
237,220
246,080
-3.60%
Backpressure
360
348
3.45%
SecurityGuard
TM
ULTRA
UHPLC Column Protection System for Any Sub-2 µm or Kinetex
®
Column
Protects with No Loss of Column Performance!
With SecurityGuard ULTRA, contaminants and microparticulates are trapped within the
guard cartridge instead of in your expensive UHPLC column. The extremely low dead vol-
ume (<0.3 L) of this unique guard design minimizes sample peak dispersion to maintain
column performance without altering your chromatography results.
Kinetex 1.7 µm column with and without the
SecurityGuard ULTRA cartridge system
For more details on the test methodology and results, contact Phenomenex.
See pp. 53 and 55 for SecurityGuard ULTRA ordering information.
0
20
40
60
80
19860
1
2
3
4
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
min
App ID 19860
App ID 19859
WITH
SecurityGuard ULTRA
WITHOUT
SecurityGuard ULTRA
Phenomenex
l
WEB: www.phenomenex.com